| Varian SpectrAA 200 Atomic Absorption Spectrophotometer with Graphite Furnace Atomization |
![]() |
The SpectrAA is a state-of-the art computer-controlled instrument, providing automation to most steps of the analysis. It is important, however, not to understand the components of the instrument to ensure successful operation and reliable results.
| The figure at left illustrates the autosampler
and graphite furnace components of the SpectrAA. Click on a letter for a brief
description of each component. The yellow arrow in the figure shows the path of light left to right from the source, through the furnace and to the detector. |
Proper alignment of the graphite furnace in the assembly is plays a major role in the quality of the results obtained in a GFAAS experiment. The assembly itself is water cooled (cooling water must be turned on prior to use) and employs an inert gas such as argon to flush material from the furnace after each measurement (argon must also be turned on prior to use). Once properly aligned, measurements are automated and can be very reproducible.
|
![]() |
C. Autosampler and Sample Carousel
The sample carousel can hold up to 50 samples, in addition to standard and blank solutions for use in an analysis. Each sample (contained in a small sample vial), is individually accessible with the sampling arm through the Varian software.
|
![]() |
To help keep the sample capillary tube clean, it is periodically flushed with a dilute solution of trace-metal grade nitric acid. A reservoir for this solution is screwed onto the bottom of the autosampler.
This plastic cover should be placed on top of the sample carousel to help minimize exposure to particulates and other possible contaminants. The cover is "keyed" so that it only fits in one orientation.